Biosensors (Sep 2024)
Microfluidic Detection Platform for Determination of Ractopamine in Food
Abstract
A novel microfluidic ractopamine (RAC) detection platform consisting of a microfluidic RAC chip and a smart analysis device is proposed for the determination of RAC concentration in meat samples. This technology utilizes gold nanoparticles (AuNPs) modified with glutamic acid (GLU) and polyethyleneimine (PEI) to measure RAC concentration in food products. When RAC is present, AuNPs aggregate through hydrogen bonding, causing noticeable changes in their optical properties, which are detected using a self-built UV–visible micro-spectrophotometer. Within the range of 5 to 80 ppb, a linear relationship exists between the absorbance ratio (A693nm/A518nm) (Y) and RAC concentration (X), expressed as Y = 0.0054X + 0.4690, with a high coefficient of determination (R2 = 0.9943). This method exhibits a detection limit of 1.0 ppb and achieves results within 3 min. The practical utility of this microfluidic assay is exemplified through the evaluation of RAC concentrations in 50 commercially available meat samples. The variance between concentrations measured using this platform and those determined via liquid chromatography–tandem mass spectrometry (LC-MS/MS) is less than 8.33%. These results underscore the viability of the microfluidic detection platform as a rapid and cost-effective solution for ensuring food safety and regulatory compliance within the livestock industry.
Keywords