Polymers (May 2024)

Smartphone-Based Rapid Quantitative Detection Platform with Imprinted Polymer for Pb (II) Detection in Real Samples

  • Flor de Liss Meza López,
  • Christian Jacinto Hernández,
  • Jaime Vega-Chacón,
  • Juan C. Tuesta,
  • Gino Picasso,
  • Sabir Khan,
  • María D. P. T. Sotomayor,
  • Rosario López

DOI
https://doi.org/10.3390/polym16111523
Journal volume & issue
Vol. 16, no. 11
p. 1523

Abstract

Read online

This paper reports the successful development and application of an efficient method for quantifying Pb2+ in aqueous samples using a smartphone-based colorimetric device with an imprinted polymer (IIP). The IIP was synthesized by modifying the previous study; using rhodizonate, 2-acrylamido-2-methylpropane sulfonic acid (AMPS), N,N′-methylenebisacrylamide (MBA), and potassium persulfate (KPS). The polymers were then characterized. An absorption study was performed to determine the optimal conditions for the smartphone-based colorimetric device processing. The device consists of a black box (10 × 10 × 10 cm), which was designed to ensure repeatability of the image acquisition. The methodology involved the use of a smartphone camera to capture images of IIP previously exposed at Pb2+ solutions with various concentrations, and color channel values were calculated (RGB, YMK HSVI). PLS multivariate regression was performed, and the optimum working range (0–10 mg L−1) was determined using seven principal components with a detection limit (LOD) of 0.215 mg L−1 and R2 = 0.998. The applicability of a colorimetric sensor in real samples showed a coefficient of variation (% RSD) of less than 9%, and inductively coupled plasma mass spectrometry (ICP–MS) was applied as the reference method. These results confirmed that the quantitation smartphone-based colorimetric sensor is a suitable analytical tool for reliable on-site Pb2+ monitoring.

Keywords