Clinical Ophthalmology (Nov 2017)

Agreement of swept-source and spectral-domain optical coherence-tomography retinal thickness measurements in neovascular age-related macular degeneration

  • Hanumunthadu D,
  • Ilginis T,
  • Balaggan KS,
  • Patel PJ

Journal volume & issue
Vol. Volume 11
pp. 2121 – 2123

Abstract

Read online

Daren Hanumunthadu,1 Tomas Ilginis,1 Kamaljit S Balaggan,1,2 Praveen J Patel1 On behalf of the DOCS team 1NIHR Biomedical Research Centre, Moorfields Eye Hospital NHS Foundation Trust and University College London Institute of Ophthalmology, London, 2Wolverhampton and Midland Counties Eye Infirmary, Wolverhampton, UKChanges in macular thickness measurements in neovascular age-related macular degeneration (nAMD) are thought to reflect disease activity, and can be used to guide treatment with antiangiogenic agents.1 Both spectral-domain optical coherence tomography (SD-OCT) and the more recently introduced swept-source optical coherence tomography (SS-OCT) have onboard segmentation algorithms that deliver automated retinal thickness measurements. Agreement between OCT derived macular thickness measurements is critical for monitoring of patients if these modalities are to be used interchangeably. This prospective cross-sectional study sought to determine the limits of agreement between SD-OCT (Spectralis; Heidelberg Engineering) and SS-OCT (DRI OCT-1 Atlantis; Topcon Medical Systems) in patients with nAMD.

Keywords