Geoscientific Instrumentation, Methods and Data Systems (Jun 2024)

Copper permalloys for fluxgate magnetometer sensors

  • B. B. Narod,
  • D. M. Miles

DOI
https://doi.org/10.5194/gi-13-131-2024
Journal volume & issue
Vol. 13
pp. 131 – 161

Abstract

Read online

Fluxgate magnetometers are commonly used to provide high-fidelity vector magnetic field measurements. The magnetic noise of the measurement is typically dominated by that intrinsic to a ferromagnetic core used to modulate (gate) the local field as part of the fluxgate sensing mechanism. A polycrystalline molybdenum–nickel–iron alloy (6.0–81.3 Mo permalloy) has been used in fluxgates since the 1970s for its low magnetic noise. Guided by previous investigations of high-permeability copper–nickel–iron alloys, we investigate alternative materials for fluxgate sensing by examining the magnetic properties and fluxgate performance of that permalloy regime in the range 28 %–45 % Cu by weight. Optimizing the alloy constituents within this regime enables us to create fluxgate cores with both lower noise and lower power consumption than equivalent cores based on the traditional molybdenum alloy. Racetrack geometry cores using six layers of ∼30 mm long foil washers consistently yield magnetic noise around 4–5 pT/Hz at 1 Hz and 6–7 pT/Hz at 0.1 Hz, meeting the 2012 1 s INTERMAGNET standard of less than 10 pT/Hz noise at 0.1 Hz.