Indian Journal of Community Medicine (Jan 2023)

A validation study on the physiological parameters recorded by COVIDBEEP - An indigenous remote health monitoring system designed for COVID-19 care in India

  • Madhuri Taranikanti,
  • Aswin Kumar Mudunuru,
  • Farheen Fatima,
  • Srinivas Maddur,
  • Rajiv Kumar Bandaru,
  • Sai Shriya Taranikanti,
  • Rohith Kumar Guntuka,
  • Aruna Kumari Yerra,
  • G Dhanunjaya,
  • A P Raju

DOI
https://doi.org/10.4103/ijcm.ijcm_80_23
Journal volume & issue
Vol. 48, no. 3
pp. 422 – 429

Abstract

Read online

Background: COVID-19 pandemic has affected mankind globally. After the three waves since March 2020, the threat continues instilling fear in the minds. Vital parameter monitoring through remote health monitoring system (RHMS) becomes critical for effective disease management and manpower safety and confidence. In a low resource setting like India, a comprehensive, wearable, and remotely operable device that is economical was required to be introduced for COVID-19 care. Present study validated the remote health monitoring device named COVIDBEEP with gold standard equipment. Materials and Methods: Six parameters, namely heart rate, SpO2, respiratory rate, temperature, blood pressure, and ECG were acquired in the supine position using the devices. Result: Analysis was performed using Graph Pad Prism. Intraclass correlation coefficients were used to measure concurrent validity. Bland–Altman graphs were plotted to know the agreement for each vital parameter. Confidence limits were set at 95%. All the parameters recorded from the devices showed a significant correlation with an “r” value between 0.5 and 0.9 with P value between 0.001 and 0.0002. Bland–Altman plots showed a minimum bias of 0.033 for heart rate and maximum of 3.5 for systolic blood pressure and respiratory rate. Conclusion: The association between the parameters recorded by the devices strengthened as the time of collection of data increased. Agreement between the two methods in 95% confidence interval was also proven to be significant for the parameters. Therefore, the indigenously developed COVIDBEEP has shown good validity in comparison to standard monitoring device.

Keywords