Journal of Oral Research (Apr 2019)

Evaluation of intensity standards of tungsten-halogen and led curing units

  • Jaime Sarmiento-Cornejo,
  • María Mackenney-Palamara

DOI
https://doi.org/10.17126/joralres.2019.017
Journal volume & issue
Vol. 8, no. 2
pp. 104 – 107

Abstract

Read online

Current evidence indicates that the minimum light intensity of photo curing units required to polymerize in a reliable way a composite resin, in increments of 2mm, is 300mW/cm2. The recent introduction of new generations of composite resin materials for large volume increments, partially contrasts with ISO 4049 (2009), calling for the use of light intensity of 1,000mW/cm2. Therefore, it is considered relevant to carry out periodic measurements of the emission intensity of light-curing units of clinical use. The aim of this study was to test the intensity [mW/cm2] of a representative sample of tungsten-halogen and LED photopolymerization units used in private and public health service in different areas of the Valparaíso Region in Chile. This was achieved through the use of dental radiometers, without considering the variables of intensity modification over time (either spontaneously, by undesirable inherent characteristics of the device, or by programs of intensity modification in time), or the density of accumulated power needed. This in vitro diagnostic test, evaluated a sample of 507 units, 107 halogen and 400 LED, for a period of around one month, using two radiometers as measuring instruments. For LED units the Bluephase Meter® radiometer, from Ivoclar-VivadentTM was used, and for halogen units we used the Coltolux® from ColténeTM. As a result, 85% of the LED and halogen units achieved the minimum requirements of intensity needed for the polymerization of conventional dental biomaterials. However, only 25% from the tested units achieved a power density of 1,000mW/cm2.

Keywords