Sensors (Apr 2023)

Validation of a Skin Calorimeter to Determine the Heat Capacity and the Thermal Resistance of the Skin

  • Pedro Jesús Rodríguez de Rivera,
  • Miriam Rodríguez de Rivera,
  • Fabiola Socorro,
  • Manuel Rodríguez de Rivera

DOI
https://doi.org/10.3390/s23094391
Journal volume & issue
Vol. 23, no. 9
p. 4391

Abstract

Read online

In vivo determination of the skin’s thermal properties is of growing interest. Several types of sensors are being designed and tested. In this field, we have developed a skin calorimeter for the determination of the heat flow, the heat capacity and the thermal resistance of the skin. The calorimeter calibration consists of the determination of the parameters of the model we have chosen to represent the behavior of the device. This model considers the heat capacity and the thermal resistance of the skin, which depend on the case (body zone, subject, physical state, etc.) and also have a strong time dependence. Therefore, this work includes a validation study with reference materials. Finally, it is concluded that the heat capacity determined is a function of the thermal penetration depth of the measurement characteristics. In the case of high thermal conductivity materials in which the thermal penetration is nearly total, the heat capacity obtained coincides with that of the reference material sample.