Journal of Pharmaceutical Analysis (Aug 2013)

Ultra-high-performance liquid chromatography for the determination of exenatide in monkey plasma by tandem quadrupole mass spectrometry

  • Jin-Feng Zhang,
  • Chun-Jie Sha,
  • Yu Sun,
  • Yun-Yun Gai,
  • Ji-Ye Sun,
  • Jiang-Bin Han,
  • Xin Shao,
  • Chun-Na Sha,
  • You-Xin Li,
  • Wan-Hui Liu

Journal volume & issue
Vol. 3, no. 4
pp. 235 – 240

Abstract

Read online

A highly sensitive ultra-high-performance liquid chromatographic–tandem mass spectrometry (UPLC/MS/MS) method was developed for the quantification of the synthetic peptide drug of exenatide in monkey plasma. Sample preparation was carried out by solid-phase extraction (SPE), and bivalirudin was used as the internal standard (IS). An excellent chromatographic separation was obtained on a reversed-phase C18 column with a gradient elution. Detection utilized a Qtrap 5500 system operated in the positive ion mode with multiple reaction monitoring (MRM). The proposed method was validated by assessing the specificity, linearity, intra- and inter-day precision and accuracy, recovery, and stability. The method resulted in a linear calibration range of 0.10–30ng/mL, extracting with only 50μL monkey plasma aliquots. The intra- and inter-day precisions (as relative standard deviation) were less than 7.5% and 9.6%, respectively. The method could be successfully utilized for the pharmacokinetic study of exenatide in monkeys following a single subcutaneous injection of Byetta. Keywords: Exenatide, UPLC/MS/MS, Quantitation, Pharmacokinetics, Plasma, Monkey