Heliyon (Aug 2022)
Simple defocus laser irradiation to suppress self-absorption in laser-induced breakdown spectroscopy (LIBS)
Abstract
This study introduces a novel and simple way to suppress the self-absorption effect in laser-induced breakdown spectroscopy (LIBS) by utilizing a defocusing laser irradiation technique. For this purpose, a Nd:YAG laser with a wavelength of 1,064 nm and repetition rate of 10 Hz with energy in the range of 10 mJ–50 mJ was used. The laser irradiation was focused by using a 150-mm-focal-length plano-convex lens onto the sample surface under defocusing of approximately –6 mm. Potassium chloride (KCl) and sodium chloride (NaCl) pellet samples were used to demonstrate this achievement. When the defocus position is adjusted to –6 mm for KCl and NaCl samples, the self-reversal in the emission lines of K I 766.4 nm, K I 769.9 nm, Na I 588.9 nm, and Na I 589.5 nm vanish. Meanwhile, the FWHM values of K I 766.4 and K I 769.9 nm are 0.29 nm and 0.23 nm, respectively, during –6 mm defocus laser irradiation, as opposed to 1.24 nm and 0.86 nm under tight focus laser irradiation. Additionally, this work demonstrates that, when the laser energy is changed between 10 and 50 mJ, no self-reversal occurs in the emission lines when –6 mm defocus laser irradiation is applied. Finally, a linear calibration curve was generated using KCl at a high concentration ranging between K concentrations from 16.6% to 29%. It should be noted that, even at such high K concentrations, the calibration curve is still linear. This means that self-absorption is almost negligible. This simple change in defocus laser irradiation will undoubtedly contribute to the suppression of the self-absorption phenomenon, which disrupts LIBS analytical results.