EPJ Web of Conferences (Jan 2020)

Design by numerical simulation of an in situ alpha spectrometer operating at ambient air pressure

  • Degrelle D.,
  • Venara J.,
  • Ben Mosbah M.,
  • Cuozzo M.,
  • Mahé C.,
  • Serrano R.

DOI
https://doi.org/10.1051/epjconf/202022506004
Journal volume & issue
Vol. 225
p. 06004

Abstract

Read online

Alpha emitters are usually identified and quantified by alpha spectrometry measurements in a vacuum chamber performed in laboratory environments. This study shows that transuranic elements can be distinguished under ambient conditions using a grid collimator. The aim of this work was to use numerical simulations with the MCNP6 code to design a grid with a resolution high enough to differentiate the same radionuclide combinations as alpha spectrometry in a vacuum chamber, namely 239Pu + 240Pu, 241Am + 238Pu and 244Cm. Results show that a compromise is required to obtain the best performances in terms of energy resolution and detection efficiency, leading to the choice of two hexagonal grid collimators. The first has a collimation height of 0.5 cm and an apothem of 1 mm. Laboratory tests on electrodeposited sources show that the target radionuclides can be identified without prior deconvolution, with an energy resolution of about 70 keV and a detection efficiency of 0.74% at incident energies of 5–6 MeV. The second grid has the same collimation height but a coarser mesh with an apothem of 2 mm. In this case, the alpha peaks are still distinguishable, but with a lower resolution of 125 keV. The detection efficiency is three times higher however.