Journal of the Serbian Chemical Society (Apr 2001)

The improvement of the detection power of a U-shaped DC plasma

  • DRAGAN MARKOVIC,
  • IVANKA HOLCLAJTNER-ANTUNOVIC,
  • MOMIR MARINKOVIC,
  • MIRJANA TRIPKOVIC

Journal volume & issue
Vol. 66, no. 4
pp. 249 – 258

Abstract

Read online

Optimization of the operating parameters of U-shpaed DC arc plasma and spectrometer parameters has been undertaken to explore the possibilities of improving its detection power. It is demonstrated, with a U-shaped arc as an example, that the limits of detection, in addition to well-defined parameters as described by Boumans and Winge, depend on the signal integration time. It is shown that with increasing integration time, the limits of detection are decreased within some limits and that the precision and concentration sensitivity are improved as well. A mathematical expression for the dependence of the detection limit on the integration time is presented. To increase the reliability of the measurement of the mentioned parameters, the working conditions were optimized for the following analytes: Ag, Al, Au, Cr, Fe, Mn, Ni, Pb, Pd, Pt, and V. The obtained limits of detection are comparable or better than those obtained by ICP for the elements studied. Itwas estimed that the possibility exists for their further improvement up to 10 times.

Keywords