Journal of Cardiovascular Magnetic Resonance (Aug 2019)

High-energy external defibrillation and transcutaneous pacing during MRI: feasibility and safety

  • Vladimir Shusterman,
  • Denice Hodgson-Zingman,
  • Daniel Thedens,
  • Xiaodong Zhu,
  • Stacy Hoffman,
  • Jessica C. Sieren,
  • Gina M. Morgan,
  • Anthony Faranesh,
  • Barry London

DOI
https://doi.org/10.1186/s12968-019-0558-z
Journal volume & issue
Vol. 21, no. 1
pp. 1 – 14

Abstract

Read online

Abstract Background Rapid application of external defibrillation, a crucial first-line therapy for ventricular fibrillation and cardiac arrest, is currently unavailable in the setting of magnetic resonance imaging (MRI), raising concerns about patient safety during MRI tests and MRI-guided procedures, particularly in patients with cardiovascular diseases. The objective of this study was to examine the feasibility and safety of defibrillation/pacing for the entire range of clinically useful shock energies inside the MRI bore and during scans, using defibrillation/pacing outside the magnet as a control. Methods Experiments were conducted using a commercial defibrillator (LIFEPAK 20, Physio-Control, Redmond, Washington, USA) with a custom high-voltage, twisted-pair cable with two mounted resonant floating radiofrequency traps to reduce emission from the defibrillator and the MRI scanner. A total of 18 high-energy (200-360 J) defibrillation experiments were conducted in six swine on a 1.5 T MRI scanner outside the magnet bore, inside the bore, and during scanning, using adult and pediatric defibrillation pads. Defibrillation was followed by cardiac pacing (with capture) in a subset of two animals. Monitored signals included: high-fidelity temperature (0.01 °C, 10 samples/sec) under the pads and 12-lead electrocardiogram (ECG) using an MRI-compatible ECG system. Results Defibrillation/pacing was successful in all experiments. Temperature was higher during defibrillation inside the bore and during scanning compared with outside the bore, but the differences were small (ΔT: 0.5 and 0.7 °C, p = 0.01 and 0.04, respectively). During scans, temperature after defibrillation tended to be higher for pediatric vs. adult pads (p = 0.08). MR-image quality (signal-to-noise ratio) decreased by ~ 10% when the defibrillator was turned on. Conclusions Our study demonstrates the feasibility and safety of in-bore defibrillation for the full range of defibrillation energies used in clinical practice, as well as of transcutaneous cardiac pacing inside the MRI bore. Methods for Improving MR-image quality in the presence of a working defibrillator require further study.

Keywords