Journal of Isotopes (Aug 2022)
Energy Spectrum to Dose Conversion Method Used in Environmental Radiation Dose Measurement
Abstract
In order to accurately measure the environmental level dose rate, based on the G(E) function method and conversion of complete spectra without deconvolution method, the energy spectrum and dose rate of X/γ ray was measured for calculating spectrum-dose conversion coefficient. And finally the conversion of energy spectrum to dose rate is realized. The residual difference between the calculated dose rate by the energy spectrum-dose method and the measured dose rate value is approximately 0, which proves that the method is self-consistent. Experiments have been conducted in 137Cs and 60Co radiation fields, and the maximum relative error of dose rate is ±10%, which proves that the G(E) function method and Conversion of complete spectra without deconvolution method are feasible for energy spectrum-dose conversion.