Analytical and Bioanalytical Chemistry Research (Jun 2016)

Determination of Rhodium(III) Ions by Flame Atomic Absorption Spectrometry after Preconcentration with Modified Magnetic Activated Carbon

  • Maryam Fayazi,
  • Masoud Ghanei-Motlagh,
  • Mohammad Ali Taher,
  • Raziyeh Fayazi

DOI
https://doi.org/10.22036/abcr.2016.14570
Journal volume & issue
Vol. 3, no. 1
pp. 87 – 99

Abstract

Read online

A new method for analysis of trace amount of Rh(III) ions by magnetic activated carbon modified with 2,3,5,6-tetra(2-pyridyl)pyrazine (MAC/TPPZ) as the magnetic sorbent has been proposed. The proposed adsorbent was found to be advantageous over conventional solid phase extraction (SPE) in terms of operational simplicity and low time-consuming. The experimental parameters affecting the extraction/preconcentration and determination of the analyte were systematically examined. In order to investigate the selectivity of this magnetic sorbent, the effect of a variety of ions on preconcentration and recovery of Rh(III) ions were also investigated. Under optimum conditions, the calibration graph was linear for the concentration range of 0.8-650 µg l-1. The limit of detection (LOD, 3Sb/m) and the relative standard deviation (RSD, n = 8, c = 50 µg l-1) were 0.1 µg l-1 and 3.6%, respectively. The maximum sorption capacity of the adsorbent for rhodium was found to be 21.6 mg g-1. The presented procedure was applied to monitoring rhodium in water and synthetic samples.

Keywords