Diagnostics (Jun 2024)
A Novel Noninvasive Screening Tool for Dry Eye Disease
Abstract
Purpose: To assess the feasibility and the diagnostic accuracy of the new tool, DEvice© (AI, Rome, Italy), for screening patients with dry eye disease (DED). Methods: This study was performed at the University Magna Græcia of Catanzaro. Enrolled patients were classified as affected by DED (group 1) or not (group 2) using an already validated tool (Keratograph 5M, Oculus, Germany), evaluating the noninvasive keratograph breakup time (NIKBUT), tear meniscus height (TMH), meibomian gland loss (MGL), and bulbar redness. All the patients were then examined by means of DEvice©, which allowed the measurement of the relative humidity (RH) and temperature of the ocular surface. Symptoms were scored using the Ocular Surface Disease Index (OSDI) questionnaire. Results: Overall, 40 patients (17 males and 23 females, mean age 38.0 ± 17.1 years) were included: of these, 20 belonged to group 1 and the remaining 20 to group 2. Using Keratograph 5M, significant differences between groups 1 and 2 were detected for NIKBUT-first (respectively, 4.97 ± 1.85 vs. 13.95 ± 4.8 s; p p = 0.0003). No statistically significant changes were detected for TMH (p = 0.565), MGL (p = 0.051), and bulbar redness (p = 0.687). Using Device©, a statistically significant higher value of RH was found in group 1 compared to group 2 (respectively, 85.93 ± 10.63 vs. 73.05 ± 12.84%; p = 0.0049). A statistically significant correlation was found between RH and OSDI (r = 0.406; p = 0.009). The value RH showed a discriminating power to detect DED with an AUC = 0.782 (standard error 0.07264; 95% CI 0.6401–0.9249; p = 0.0022). Conclusions: The DEvice© can effectively discriminate DED patients from healthy subjects. The parameter RH showed good sensitivity, making this tool ideal for a fast and noninvasive DED screening.
Keywords