MethodsX (Jun 2024)

A green chemical analysis of iron (III) in water sample using a smart phone

  • Jalal Hassan,
  • Safdar Mehdizadeh Shermeh,
  • Mohammad Kazem Koohi,
  • Ali Pourshaban-Shahrestani

Journal volume & issue
Vol. 12
p. 102724

Abstract

Read online

In this study, a smartphone was used as the photometer for determination of iron (III) by forming a complex with thiocyanate. After color formation at concentrations of, image capture with mobile phone, signal analysis of each sample was performed by the application and converted to the absorption number. The calibration curve was completely linear in the range of 10 to 80 mg L–1 used and the linear coefficient was better than 0.9833. The limits of detection (LOD) and quantification (LOQ) were 0.1 and 0.3 mg L–1, respectively. Finally, this method was successfully used to measure iron in real samples. • A smartphone was used for the determination of iron (III), showcasing its potential in color compound analysis. • The method demonstrated superior performance in terms of calibration curve range and measurement speed compared to traditional atomic absorption devices. • The technique was successfully applied in the measurement of iron in real samples, indicating its practical applicability.