Journal of Arrhythmia (Feb 2017)

Inappropriate shocks from a subcutaneous implantable cardioverter-defibrillator due to oversensing during periods of rate-related bundle branch block

  • Maria João Sousa, MD,
  • Tim Betts, MD

DOI
https://doi.org/10.1016/j.joa.2016.05.006
Journal volume & issue
Vol. 33, no. 1
pp. 73 – 75

Abstract

Read online

The subcutaneous implantable cardioverter-defibrillator (S-ICD) is a novel technology with proven efficacy in sudden cardiac death prevention; however, there is a lack of long-term safety data. We describe the case of a 55-year-old female patient implanted with an S-ICD due to idiopathic ventricular fibrillation, who subsequently presented with inappropriate shocks leading to ventricular fibrillation that was successfully terminated by another shock. Inappropriate shocks were due to intermittent T wave oversensing during periods of rate-dependent right bundle branch block. Assessment of the S-ICD electrograms during an exercise test allowed successful reprogramming of the device׳s sensing vector with no further events.