Journal of Medical Physics (Jan 2017)

A phantom study on fetal dose reducing factors in pregnant patients with breast cancer during radiotherapy treatment

  • Akın Ogretici,
  • Aydın Çakır,
  • Uğur Akbaş,
  • Canan Köksal,
  • Ümmühan Kalafat,
  • Makbule Tambaş,
  • Hatice Bilge

DOI
https://doi.org/10.4103/jmp.JMP_133_16
Journal volume & issue
Vol. 42, no. 3
pp. 128 – 132

Abstract

Read online

Purpose: This study aims to investigate the factors that reduce fetal dose in pregnant patients with breast cancer throughout their radiation treatment. Two main factors in a standard radiation oncology center are considered as the treatment planning systems (TPSs) and simple shielding for intensity modulated radiation therapy technique. Materials and Methods: TPS factor was evaluated with two different planning algorithms: Anisotropic analytical algorithm and Acuros XB (external beam). To evaluate the shielding factor, a standard radiological purpose lead apron was chosen. For both studies, thermoluminescence dosimeters were used to measure the point dose, and an Alderson RANDO-phantom was used to simulate a female pregnant patient in this study. Thirteen measurement points were chosen in the 32nd slice of the phantom to cover all possible locations of a fetus up to 8th week of gestation. Results: The results show that both of the TPS algorithms are incapable of calculating the fetal doses, therefore, unable to reduce them at the planning stage. Shielding with a standard lead apron, however, showed a slight radiation protection (about 4.7%) to the fetus decreasing the mean fetal dose from 84.8 mGy to 80.8 mGy, which cannot be disregarded in case of fetal irradiation. Conclusions: Using a lead apron for shielding the abdominal region of a pregnant patient during breast irradiation showed a minor advantage; however, its possible side effects (i.e., increased scattered radiation and skin dose) should also be investigated further to solidify its benefits.

Keywords