Antibiotics (Jun 2023)

Performance Evaluation of BD Phoenix and MicroScan WalkAway Plus for Determination of Fosfomycin Susceptibility in <i>Enterobacterales</i>

  • Alessandro Bondi,
  • Antonio Curtoni,
  • Marco Peradotto,
  • Elisa Zanotto,
  • Matteo Boattini,
  • Gabriele Bianco,
  • Marco Iannaccone,
  • Anna Maria Barbui,
  • Rossana Cavallo,
  • Cristina Costa

DOI
https://doi.org/10.3390/antibiotics12071106
Journal volume & issue
Vol. 12, no. 7
p. 1106

Abstract

Read online

Background: Fosfomycin is an old bactericidal drug that has gained increasing interest in the last decade for its potential use in multi-drug resistant gram-negative infections. However, evidence on fosfomycin susceptibility testing reports a poor correlation between commercial methods vs. reference agar dilution (AD) for Enterobacterales (EB). The study aimed at assessing the performance of two automated systems for the determination of fosfomycin susceptibility in EB clinical isolates. Methods: Fosfomycin susceptibility testing results of two collections of 100 non-duplicate clinical EB strains obtained using two different platforms (BD Phoenix and MicroScan WalkAway Plus) were compared with those obtained by AD. Categorical agreement (CA), major error (ME) and very major error (VME) rates were calculated. Results: BD Phoenix exhibited a 6.9% rate of false-resistant results and achieved a CA of 69%, whereas MicroScan WalkAway Plus achieved 3.7% of false-resistant results and 72% of CA. Both automated systems showed poor detection of resistant isolates, with 49.1% and 56.2% of false-susceptible results for BD Phoenix and Microscan WalkAway Plus, respectively. Conclusions: Overall, agar dilution remains the most suitable method for routine laboratory antimicrobial susceptibility testing of fosfomycin on Enterobacterales strains, given the poor performance of automated systems. The application of both automated systems, in the clinical laboratories reporting of fosfomycin, should be reviewed in light of the accuracy results falling below the acceptable threshold.

Keywords