Scientific Reports (Jun 2024)

Hemodynamics with mechanical circulatory support devices using a cardiogenic shock model

  • Kazuyuki Yahagi,
  • Gohki Nishimura,
  • Kei Kuramoto,
  • Yusuke Tsuboko,
  • Kiyotaka Iwasaki

DOI
https://doi.org/10.1038/s41598-024-64721-1
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 11

Abstract

Read online

Abstract Mechanical circulatory support (MCS) devices, including veno-arterial extracorporeal membrane oxygenation (VA-ECMO) and Impella, have been widely used for patients with cardiogenic shock (CS). However, hemodynamics with each device and combination therapy is not thoroughly understood. We aimed to elucidate the hemodynamics with MCS using a pulsatile flow model. Hemodynamics with Impella CP, VA-ECMO, and a combination of Impella CP and VA-ECMO were assessed based on the pressure and flow under support with each device and the pressure–volume loop of the ventricle model. The Impella CP device with CS status resulted in an increase in aortic pressure and a decrease in end-diastolic volume and end-diastolic pressure (EDP). VA-ECMO support resulted in increased afterload, leading to a significant increase in aortic pressure with an increase in end-systolic volume and EDP and decreasing venous reservoir pressure. The combination of Impella CP and VA-ECMO led to left ventricular unloading, regardless of increase in afterload. Hemodynamic support with Impella and VA-ECMO should be a promising combination for patients with severe CS.

Keywords