CVIR Endovascular (Mar 2020)

Flow diversion in challenging vascular anatomies: the use of low profile stent retrievers for safe and accurate positioning of the microcatheter

  • Ulf Quäschling,
  • Monika Kläver,
  • Cindy Richter,
  • Gordian Hamerla,
  • Simone Mucha,
  • Cordula Scherlach,
  • Jens Maybaum,
  • Karl-Titus Hoffmann,
  • Stefan Schob

DOI
https://doi.org/10.1186/s42155-020-00106-5
Journal volume & issue
Vol. 3, no. 1
pp. 1 – 7

Abstract

Read online

Abstract Background Hemodynamic therapy with Flow-Diverters has become a fundamental option for treatment of cerebral aneurysms. A major obstacle of Flow-Diverters is the comparatively stiff microcatheter required for implantation. Consequentially, maneuverability is limited and primary catheterization of peripheral targets may be difficult or even futile in challenging vascular anatomies. To overcome this, a highly navigable microcatheter must be used to attain the desired vascular segment, followed by a hardly controllable exchange-maneuver via a long microwire, involving a high risk for wire-perforation. Our study aimed to investigate the value of low-profile stent-retrievers as a railway for introduction of the required microcatheter, which allows to maintain a stable endovascular position and reduce the risk for procedural vessel injury. Methods 14cases (8females, mean-age 59y) of Flow-Diverter-Implantation requiring the use of a low-profile stent-retriever were reviewed. All cases featured a challenging vascular anatomy. After micro-catheterization of the desired segment, the stent-retriever was carefully deployed as an anchor in a secure, distal location. In all cases a pREset/LITE-stent-retriever was used for introduction of the equipment required for implantation. Results In all cases the anchoring-maneuver was performed without technical complications. The stent-retrievers maintained a stable position after deployment in all situations. No potential traumatic sudden movements of the microcatheter occurred. No procedure-related perforations, dissections or vasospasms were observable during the interventions or their aftermath. Conclusions In our experience the stent-retriever-anchoring-maneuver represents a potentially essential and safe amendment for flow diverter treatment in technically challenging situations.

Keywords