Journal of Isotopes (Feb 2021)

Preparation of Standard Plane Source of 210Po

  • ZHU Heng;CHEN Xilin;ZHU Baoji;DIAO Lijun;YAO Shunhe;XIN Yu;GUO Xiaoqing;YAO Yanling;LYU Xiaoxia

Journal volume & issue
Vol. 34, no. 1
pp. 61 – 65

Abstract

Read online

A method of spontaneous deposition of 210Po on a silver disc was used to prepare the standard plane source in a 210Pb solution. The prepared source was measured by an alpha spectrometer. There was a unique alpha peak of 5.304 MeV with a full width at half maximum of 23.92 keV in the measured spectrum. A plateau length of about 1000 V was given in the plateau curve measued with a multi-wire proportional counter. The determined surface emission rate was 2.565 ×104 min-1 at JAN. 3RD , 2019, with the extended uncertainty of 2.7% (k=2). Some MPC9604 gas-flow low background gross alpha and gross beta counters were used to test the source according to the verification regulation JJG 1100-2014. All the α detection efficiencies were not less than 77%, the repeatability within 1.5% and the α interference to β channel not greater than 2.5%. It’s showed that the prepared source was in good quality and performance to meet the calibration of gas-flow low background gross alpha and gross beta counters.