BMC Ophthalmology (Jul 2024)

Comparison of axial length and anterior segment parameters of patients with myopia measured using 2 fourier-domain optical coherent biometry devices

  • Bingqing Sun,
  • Yuhao Ye,
  • Jing Zhao,
  • Xingtao Zhou,
  • Lingling Niu

DOI
https://doi.org/10.1186/s12886-024-03546-y
Journal volume & issue
Vol. 24, no. 1
pp. 1 – 8

Abstract

Read online

Abstract Background This study assessed the agreement of ocular parameters of patients with myopia measured using Colombo intraocular lens (IOL) 2 and IOLMaster 700. Methods Eighty patients (male, 22; average age, 29.14 ± 7.36 years) with myopia (159 eyes) were included in this study in May 2023. The participants’ axial length (AXL), central corneal thickness (CCT), lens thickness (LT), white-to-white distance (WTW), front flat (K1), steep (K2), mean (Km) corneal keratometry, astigmatism (Astig), J0 vector, and J45 vector were measured using the IOLMaster 700 and Colombo IOL 2. The measurements from both devices were compared using the generalized estimating equation, correlation analysis, and Bland-Altman plots. Results With the Colombo IOL 2, lower values for K2 and J0 (odds ratio [OR] = 0.587, p = 0.033; OR = 0.779, p < 0.0001, respectively), and larger values for WTW, Astig, and J45 (OR = 1.277, OR = 1.482, OR = 1.1, all p < 0.0001) were obtained. All ocular measurements by both instruments showed positive correlations, with AXL demonstrating the strongest correlation (r = 0.9996, p < 0.0001). The intraclass correlation coefficients for AXL and CCT measured by both instruments was 0.999 and 0.988 (both p < 0.0001), and Bland-Altman plot showed 95% limits of agreement (LoA) of -0.078 to 0.11 mm and − 9.989 to 13.486 μm, respectively. The maximum absolute 95% LoA for LT, WTW, K1, K2, and J0 were relatively high, achieving 0.829 mm, 0.717 mm, 0.983 D, 0.948 D, and 0.632 D, respectively. Conclusions In young patients with myopia, CCT and AXL measurements obtained with the Colombo IOL 2 and IOLMaster 700 were comparable. However, WTW, LT, corneal refractive power, and astigmatism values could not be used interchangeably in clinical practice.

Keywords