Performance of the Intermittently Scanned Continuous Glucose Monitoring (isCGM) System during a High Oral Glucose Challenge in Adults with Type 1 Diabetes—A Prospective Secondary Outcome Analysis

  • Othmar Moser,
  • Norbert Tripolt,
  • Peter Pferschy,
  • Anna Obermayer,
  • Harald Kojzar,
  • Alexander Mueller,
  • Hakan Yildirim,
  • Caren Sourij,
  • Max Eckstein,
  • Harald Sourij

DOI
https://doi.org/10.3390/bios11010022
Journal volume & issue
Vol. 11, no. 1
p. 22

Abstract

Read online

To assess intermittently scanned continuous glucose monitoring (isCGM) performance for different rates of change in plasma glucose (RCPG) during glycemic challenges in type 1 diabetes (T1D). Nineteen people with T1D (7 females; age 35 ± 11 years; HbA1c 7.3 ± 0.6% (56 ± 7 mmol/mol)) performing two glycemic challenges (OGTT) were included. During OGTTs, plasma glucose was compared against sensor glucose for timepoints 0 min (pre-OGTT), +15 min, +30 min, +60 min, +120 min, +180 min, and +240 min by means of median absolute (relative) difference (MARD and MAD) and Clarke Error Grid (CEG), then was stratified for RCPG and glycemic ranges. Overall, MARD was 8.3% (4.0–14.8) during hypoglycemia level 1 18.8% (15.8–22.0), euglycemia 9.5% (4.3–15.1), hyperglycemia level 1 9.4% (4.0–17.2), and hyperglycemia level 2 7.1% (3.3–11.9). The MARD was associated with the RCPG (p p = 0.014). Overall, CEG resulted in 88% (212 values) of comparison points in zone A, 12% (29 values) in zone B, and 0.4% (1 value) in zone D. The isCGM system was accurate during OGTTs. Its performance was dependent on the RCPG and showed an overestimation of the actual reference glucose during hypoglycemia.