Frontiers in Oncology (Jun 2023)

Evaluation of artificial intelligent breast ultrasound on lesion detection and characterization compared with hand-held ultrasound in asymptomatic women

  • Bin Xu,
  • Bin Xu,
  • Weidong Luo,
  • Weidong Luo,
  • Xin Chen,
  • Xin Chen,
  • Yiping Jia,
  • Yiping Jia,
  • Mengyuan Wang,
  • Mengyuan Wang,
  • Lulu Tian,
  • Lulu Tian,
  • Yi Liu,
  • Yi Liu,
  • Bowen Lei,
  • Bowen Lei,
  • Jiayuan Li,
  • Jiayuan Li

DOI
https://doi.org/10.3389/fonc.2023.1207260
Journal volume & issue
Vol. 13

Abstract

Read online

IntroductionTo compare the accuracy of Artificial Intelligent Breast Ultrasound (AIBUS) with hand-held breast ultrasound (HHUS) in asymptomatic women and to offer recommendations for screening in regions with limited medical resources.Methods852 participants who underwent both HHUS and AIBUS were enrolled between December 2020 and June 2021. Two radiologists, who were unaware of the HHUS results, reviewed the AIBUS data and scored the image quality on a separate workstation. Breast imaging reporting and data system (BI-RADS) final recall assessment, breast density category, quantified lesion features, and examination time were evaluated for both devices. The statistical analysis included McNemar’s test, paired t-test, and Wilcoxon test. The kappa coefficient and consistency rate were calculated in different subgroups.ResultsSubjective satisfaction with AIBUS image quality reached 70%. Moderate agreements were found between AIBUS with good quality images and HHUS for the BI-RADS final recall assessment (κ = 0.47, consistency rate = 73.9%) and breast density category (κ = 0.50, consistency rate = 74.8%). The lesions measured by AIBUS were statistically smaller and deeper than those measured by HHUS (P < 0.001), though they were not significant in clinical diagnosis (all < 3 mm). The total time required for the AIBUS examination and image interpretation was 1.03 (95% CI (0.57, 1.50)) minutes shorter than that of HHUS per case.ConclusionModerate agreement was obtained for the description of the BI-RADS final recall assessment and breast density category. With image quality comparable to that of HHUS, AIBUS was superior for the efficiency of primary screening.

Keywords