International Journal of Neonatal Screening (Oct 2022)
Evaluation of a New Laboratory Protocol for Newborn Screening for Congenital Adrenal Hyperplasia in New Zealand
Abstract
Between 2005 and 2021, 49 cases of classical congenital adrenal hyperplasia were diagnosed in New Zealand, 39 were detected in newborns and 10 were not detected by screening. Currently, for every case of CAH detected by screening, 10 false-positive tests are encountered. Second-tier liquid chromatography-tandem mass spectrometry (LCMSMS) has the potential to improve screening sensitivity and specificity. A new laboratory protocol for newborn screening for CAH was evaluated. Birthweight-adjusted thresholds for first- and second-tier 17-hydroxyprogesterone, second-tier 21-deoxycortisol and a steroid ratio were applied to 4 years of newborn screening data. The study was enriched with 35 newborn screening specimens from confirmed CAH cases. Newborn screening was conducted on 232,542 babies, and 11 cases of classical CAH were detected between 2018 and 2021. There were 98 false-positive tests (specificity 99.96%, PPV = 10.1%) using the existing protocol. Applying the new protocol, the same 11 cases were detected, and there were 13 false-positive tests (sensitivity > 99.99%, PPV = 45.8%, (X2 test p 2 test p = 0.1338). Implementation of LCMSMS as a second-tier test will improve newborn screening for classical CAH in New Zealand.