CJC Open (Jun 2022)

Transcatheter Aortic Valve Implantation in a Failed Perceval Sutureless Valve, Complicated by Aortic Annular RuptureNovel Teaching Points

  • Robert T. Kay, MB, BCh, BAO,
  • Hatem Linjawi, MBBS,
  • Craig Butler, MD, MSc,
  • Anoop Mathew, MBBS,
  • Isabelle Vonder Muhll, MD,
  • Sayra Khandekar, MD,
  • Benjamin D. Tyrrell, MD,
  • Jeevan Nagendran, MD, PhD,
  • Dylan Taylor, MD,
  • Robert C. Welsh, MD

Journal volume & issue
Vol. 4, no. 6
pp. 577 – 580

Abstract

Read online

As the use of surgically implanted sutureless aortic valves has increased over the past decade, we expect to encounter their failure increasingly in coming years. We describe a case of Perceval aortic valve failure with stent infolding and severe stenosis. This condition was treated with valve-in-valve transcatheter aortic valve implantation and complicated by aortic annular rupture at the site of infolding. This case is important because it outlines the limited experience with valve-in-valve transcatheter aortic valve implantation to treat failed sutureless valves and identifies sutureless valve infolding as a potential risk for annular rupture. Résumé: Puisque l’implantation valvulaire aortique sans suture s’est accrue au cours de la dernière décennie, nous nous attendons à rencontrer de plus en plus de défaillances de valves dans les années à venir. Nous décrivons un cas de défaillance de la valve aortique Perceval avec pliage de l’endoprothèse et sténose grave. Le traitement qui consistait en l’implantation valvulaire aortique de type valve-in-valve par cathéter a été compliqué par la rupture de l’anneau aortique au site du pliage. Il s’agit d’un cas important puisqu’il décrit le peu d’expérience en matière d’implantation valvulaire aortique de type valve-in-valve par cathéter dans le traitement des valves sans suture défectueuses et établit que le pliage d’une valve sans suture expose à un risque de rupture de l’anneau.