Scientific Reports (Jul 2024)

Development of implantable electrode based on bioresorbable Mg alloy for tissue welding application

  • Lin Mao,
  • Zhengyi Han,
  • Xupo Xing,
  • Zhongxin Hu,
  • Langlang She,
  • Chengli Song

DOI
https://doi.org/10.1038/s41598-024-67077-8
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 13

Abstract

Read online

Abstract An implantable electrode based on bioresorbable Mg-Nd-Zn-Zr alloy was developed for next-generation radiofrequency (RF) tissue welding application, aiming to reduce thermal damage and enhance anastomotic strength. The Mg alloy electrode was designed with different structural features of cylindrical surface (CS) and continuous long ring (LR) in the welding area, and the electrothermal simulations were studied by finite element analysis (FEA). Meanwhile, the temperature variation during tissue welding was monitored and the anastomotic strength of welded tissue was assessed by measuring the avulsion force and burst pressure. FEA results showed that the mean temperature in the welding area and the proportion of necrotic tissue were significantly reduced when applying an alternating current of 110 V for 10 s to the LR electrode. In the experiment of tissue welding ex vivo, the maximum and mean temperatures of tissues welded by the LR electrode were also significantly reduced and the anastomotic strength of welded tissue could be obviously improved. Overall, an ideal welding temperature and anastomotic strength which meet the clinical requirement can be obtained after applying the LR electrode, suggesting that Mg-Nd-Zn-Zr alloy with optimal structure design shows great potential to develop implantable electrode for next-generation RF tissue welding application.