Scientific Reports (Oct 2023)

Validation of quantitative loop-mediated isothermal amplification assay using a fluorescent distance-based paper device for detection of Escherichia coli in urine

  • Natkrittaya Saengsawang,
  • Panthita Ruang-areerate,
  • Nuanlaong Kaeothaisong,
  • Saovanee Leelayoova,
  • Mathirut Mungthin,
  • Piraporn Juntanawiwat,
  • Patomroek Hanyanunt,
  • Patsanun Potisuwan,
  • Piyanate Kesakomol,
  • Pornphan Butsararattanagomen,
  • Pattarawadee Wichaiwong,
  • Wijitar Dungchai,
  • Toon Ruang-areerate

DOI
https://doi.org/10.1038/s41598-023-46001-6
Journal volume & issue
Vol. 13, no. 1
pp. 1 – 10

Abstract

Read online

Abstract Uropathogenic Escherichia coli (UPEC) causes up to 90% of urinary tract infections (UTI) which is more prevalent among females than males. In urine, patients with symptomatic UTI usually have a high concentration of bacterial infection, ≥ 105 colony-forming units (CFU) per mL, in which the culture method is regularly the gold standard diagnosis. In this study, a simple and inexpensive distance-based paper device (dPAD) combined with the fluorescent closed tube LAMP assay was validated for simultaneously screening and semi-quantifying the infection level of E. coli in 440 urine samples of patients with UTI. The dPAD could measure the LAMP amplicons and semi-quantify the levels of E. coli infection in heavy (≥ 104 CFU/mL), light (≤ 103 CFU/mL) and no infection. The sensitivity and specificity had reliable performances, achieving as high as 100 and 92.7%, respectively. The one step LAMP assay could be performed within 3 h, which was 7.5 times faster than the culture method. To empower early UTI diagnosis and fast treatment, this inexpensive dPAD tool combined with the fluorescent closed tube LAMP assay is simple, reliably fast and practically portable for point-of-care settings, particularly in resource-limited areas, which can be set up in all levels of healthcare facilities.