Diagnostics (May 2022)

Repeatability of Contrast-Enhanced Ultrasound to Determine Renal Cortical Perfusion

  • Shatha J. Almushayt,
  • Alisa Pham,
  • Bethan E. Phillips,
  • John P. Williams,
  • Maarten W. Taal,
  • Nicholas M. Selby

DOI
https://doi.org/10.3390/diagnostics12051293
Journal volume & issue
Vol. 12, no. 5
p. 1293

Abstract

Read online

Alterations in renal perfusion play a major role in the pathogenesis of renal diseases. Renal contrast-enhanced ultrasound (CEUS) is increasingly applied to quantify renal cortical perfusion and to assess its change over time, but comprehensive assessment of the technique’s repeatability is lacking. Ten adults attended two renal CEUS scans within 14 days. In each session, five destruction/reperfusion sequences were captured. One-phase association was performed to derive the following parameters: acoustic index (AI), mean transit time (mTT), perfusion index (PI), and wash-in rate (WiR). Intra-individual and inter-operator (image analysis) repeatability for the perfusion variables were assessed using intra-class correlation (ICC), with the agreement assessed using a Bland–Altman analysis. The 10 adults had a median (IQR) age of 39 years (30–46). Good intra-individual repeatability was found for mTT (ICC: 0.71) and PI (ICC: 0.65). Lower repeatability was found for AI (ICC: 0.50) and WiR (ICC: 0.56). The correlation between the two operators was excellent for all variables: the ICCs were 0.99 for PI, 0.98 for AI, 0.87 for mTT, and 0.83 for WiR. The Bland–Altman analysis showed that the mean biases (± SD) between the two operators were 0.03 ± 0.16 for mTT, 0.005 ± 0.09 for PI, 0.04 ± 0.19 for AI, and −0.02 ± 0.11 for WiR.