Türk Kardiyoloji Derneği Arşivi (Feb 2016)

Inappropriate shock and battery switching to 'End of Life' in a patient with biventricular ICD during magnetic resonance imaging

  • İlyas Atar,
  • Uğur Bal,
  • Çağatay Ertan,
  • Bülent Özin,
  • Haldun Müderrisoğlu

DOI
https://doi.org/10.5543/tkda.2015.68624
Journal volume & issue
Vol. 44, no. 1
pp. 79 – 81

Abstract

Read online

Presence of a cardiac pacemaker or implantable cardioverter defibrillator (ICD) is a relative contraindication to magnetic resonance imaging (MRI). Biventricular ICDs are often used in the treatment of advanced heart failure; however, reports on experience with biventricular ICDs are lacking in the literature. In this case report, we describe a pacemaker-dependent patient with a biventricular ICD on whom an MRI of the lumbar spine was performed without having realized the presence of the ICD.

Keywords