Iranian Journal of Medical Physics (Sep 2019)

In Vivo Dosimetry Using a Flat Surface Sun Nuclear Corporation Diode in 60co Beams for Some Radiotherapy Treatments in Ghana

  • Eric Manson,
  • Vivian Atuwo‑Ampoh,
  • Samuel Tagoe,
  • Emmanuel Fiagbedzi,
  • Eric Addison,
  • Cyril Schandorf

DOI
https://doi.org/10.22038/ijmp.2018.29705.1324
Journal volume & issue
Vol. 16, no. 5
pp. 329 – 335

Abstract

Read online

Introduction: One of the useful standard quality assurance techniques in radiation therapy is monitoring entrance doses in in-vivo dosimetry. An overall tolerance limit of 5% of the absorbed radiation dose has been recommended by the International Commission of Radiological Units. The implementation of an in vivo dosimetry still remains as a challenge to clinical medical physicists. As a result, the practice of constant monitoring of patients undergoing radiation therapy in most of the radiotherapy departments in Africa has not been given much attention. The study aimed at the evaluation of in-vivo entrance dosimetry using diodes to verify the accuracy of the radiation delivered to patients, compared to prescribed doses. Material and Methods: In this paper, a protocol for in vivo dosimetry using a two flat surface Sun Nuclear Corporation diode in a radiotherapy department has been implemented in equinox Cobalt 60 beams. A water phantom calibrated was performed using the International Atomic Energy Agency standards (TRS 398). Calibration coefficients were determined with diodes using a Perspex phantom to derive correction factors. A total number of 137 patients’ doses were measured with the diodes during the treatment of 4 different sites. Results: The average deviation between the measured and expected entrance dose performed by the phantom studies was 5% (0.34±1.8%) in almost all cases. Conclusion: The developed protocol in this study indicates that in vivo dosimetry using silicon diodes is reliable, which can be adopted as a universal quality assurance tool in the radiotherapy departments. Moreover, measurements with diodes can be acquired online which produces an instant readout and is relatively cheaper as compared to the ion chamber.

Keywords