Scientific Reports (Nov 2024)
Combined influence of quantum iterative reconstruction level and kernel sharpness on image quality in photon counting CT angiography of the upper leg
Abstract
Abstract Aim was to evaluate the influence of different quantum iterative reconstruction (QIR) levels on the image quality of femoral photon-counting CT angiographies (PCD-CTA). Ultra-high resolution PCD-CTA were obtained from both extremities of five extracorporeally-perfused cadavers using constant tube voltage and maximum radiation dose (71.2 ± 11.0 mGy). Images were reconstructed with three kernels (Bv48, Bv60, Bv76) and the four available levels of QIR. Signal attenuation in the arterial lumen, muscle, and fat were measured. Contrast-to-noise ratios (CNR) and blurring scores were calculated for objective assessment. Six radiologists evaluated the subjective image quality using a pairwise comparison tool. Higher QIR level resulted in a decisive image noise reduction, especially with sharper convolution kernels (Bv60: Q1 11.5 ± 6.3 HU vs. Q4 8.4 ± 2.6 HU; p < 0.001). Largest improvement of CNR was recorded with ultra-sharp reconstructions (Bv76: Q1 20.2 ± 4.4 vs. Q4 28.0 ± 3.5; p < 0.001). Blurring decreased with higher QIR levels for soft Bv48, remained constant for medium Bv60, and increased for sharp Bv76 reconstructions. Subjective QIR level preference varied kernel depending, preferred combinations were: Bv48/Q4, Bv60/Q2, Bv76/Q3. Interrater agreement was excellent. Sharp kernels benefited most from noise reduction of higher QIR levels in lower extremity PCD-CTA. In sum, QIR level 3 provided the best objective and subjective image quality results.