The Scientific World Journal (Jan 2013)

Delivery Parameter Variations and Early Clinical Outcomes of Volumetric Modulated Arc Therapy for 31 Prostate Cancer Patients: An Intercomparison of Three Treatment Planning Systems

  • Shinichi Tsutsumi,
  • Masako N. Hosono,
  • Daisaku Tatsumi,
  • Yoshitaka Miki,
  • Yutaka Masuoka,
  • Ryo Ogino,
  • Kentaro Ishii,
  • Yasuhiko Shimatani,
  • Yukio Miki

DOI
https://doi.org/10.1155/2013/289809
Journal volume & issue
Vol. 2013

Abstract

Read online

We created volumetric modulated arc therapy (VMAT) plans for 31 prostate cancer patients using one of three treatment planning systems (TPSs)—ERGO++, Monaco, or Pinnacle—and then treated those patients. A dose of 74 Gy was prescribed to the planning target volume (PTV). The rectum, bladder, and femur were chosen as organs at risk (OARs) with specified dose-volume constraints. Dose volume histograms (DVHs), the mean dose rate, the beam-on time, and early treatment outcomes were evaluated and compared. The DVHs calculated for the three TPSs were comparable. The mean dose rates and beam-on times for Ergo++, Monaco, and SmartArc were, respectively, 174.3 ± 17.7, 149.7 ± 8.4, and 185.8 ± 15.6 MU/min and 132.7 ± 8.4, 217.6 ± 13.1, and 127.5 ± 27.1 sec. During a follow-up period of 486.2 ± 289.9 days, local recurrence was not observed, but distant metastasis was observed in a single patient. Adverse events of grade 3 to grade 4 were not observed. The mean dose rate for Monaco was significantly lower than that for ERGO++ and SmartArc (P<0.0001), and the beam-on time for Monaco was significantly longer than that for ERGO++ and SmartArc (P<0.0001). Each TPS was successfully used for prostate VMAT planning without significant differences in early clinical outcomes despite significant TPS-specific delivery parameter variations.