Journal of Pure and Applied Microbiology (Dec 2018)

Comparison of Two Detection Methods for Fragment Ions of Escherichia coli Carbon Center Metabolites

  • Bin Rui,
  • Xiaotao Wei,
  • Long Chen,
  • Yun An,
  • Han Wen

DOI
https://doi.org/10.22207/JPAM.12.4.05
Journal volume & issue
Vol. 12, no. 4
pp. 1729 – 1736

Abstract

Read online

Mass spectrometry is an analytical technique for molecules detection after separating the particle motion (charged molecules, atoms and fragment ions) according to their mass-to-charge ratio, under the action of an electric field and a magnetic field. In this study, ten kinds of fragment ions of metabolites in Escherichia coli were detected by ion trap liquid mass instrument and triple quadrupole mass chromatograph. Two different detection methods were then compared, to provide a foundation for later metabolomic research. This experiment detects metabolites with multistage mass spectrometry by electrospray ion trap mass spectrometry and triple quadrupole mass spectrometry, when scanning at anion mode using Agilent SB-C18 (250mm×4.6mm×5ìm) chromatographic column and using ammonium acetate and acetonitrile as the mobile phase. The result showed that both methods can act as multistage mass spectrometry, while triple quadrupole mass spectrometry enables to produce finer ions under low collision energy. Triple quadrupole mass spectrometry is a superior method to acquire a clear peak in detecting fragment ions of metabolites.