Poultry Science (Jul 2020)

Research Note: Comparison of chicken blood chemistry and electrolyte parameters between the portable i-STAT1 clinical analyzer and VetScan VS2 serum biochemistry panel using Hy-Line commercial white-egg laying hens

  • Z.C. Sauer,
  • K. Taylor,
  • A. Wolc,
  • A. Viall,
  • J.E. Fulton,
  • P. Settar,
  • I. Rubinoff,
  • T. Schaal,
  • Y. Sato

Journal volume & issue
Vol. 99, no. 7
pp. 3487 – 3490

Abstract

Read online

The i-STAT1 clinical analyzer has become an increasingly popular tool in clinical production animal medicine as it can provide pen-side results in a cost effective and timely manner when compared to standard benchtop serum biochemistry blood gas and chemistry analyses. This study compares the results of the portable Abbott i-STAT1 analyzer and the Abaxis VetScan VS2 for glucose (Glu, mg/dL), ionized Ca (mmol/L), Na (mmol/L), and K (mmol/L) values. Three genetically distinct commercial varieties (CV) of Hy-Line white-egg laying hens are used in this study (Hy-Line W-36, Hy-Line W-80, and Hy-Line W-80+). Thirty blood samples (n = 10 per CV) were obtained in the production house from the brachial vein and concurrently analyzed by the i-STAT1 portable device. Serum from 22 of these same samples was analyzed via VetScan VS2, a benchtop serum clinical biochemistry analyzer, using VetScan Avian/Reptilian Profile Plus reagent rotors. A paired T-test was used to test for statistical differences in means between the 2 instruments for each of the parameters. Parameters with significant mean differences were then subject to correlation and regression analysis to further evaluate relationships between the results from the 2 methods. Significant differences between means were found for Glu, Na, and K levels. Ca levels were found to be not directly comparable by the 2 analysis instruments. This comparison elucidates the importance of clinical analyzer validations when applying different strategies of diagnostic medicine in poultry.

Keywords