Guoji Yanke Zazhi (Mar 2014)

Comparison of central corneal thickness measurements after LASEK by three instruments

  • Xiao-Juan Peng,
  • Ye Tian,
  • Si-Ying Liu,
  • Ji-Sheng Chen

DOI
https://doi.org/10.3980/j.issn.1672-5123.2014.03.26
Journal volume & issue
Vol. 14, no. 3
pp. 477 – 479

Abstract

Read online

AIM: To compare the differences and the relationships of central corneal thickness(CCT)measurements among optical coherence tomography(OCT), Tomey pachymetry and ultrasound pachymetry after laser assisted subepithelial keratomileusis(LASEK). METHODS:Forty-nine patients(96 eyes)with myopia after 6mo of LASEK were assigned to measure the CCT by OCT, Tomey pachymetry and ultrasound pachymetry. Paired t-test and Pearson correlation analysis was used for statistical analysis. RESULTS: The mean CCT measured by OCT, Tomey pachymetry and ultrasound pachymetry was(475.15±33.67μm),(447.80±34.11μm),(465.18±34.23μm), respectively. Surgery reserved mean CCT was(431.22±35.32)μm. There were statistically differences between two comparative measured results(PPCONCLUSION: The CCT measurements of three instruments were not interchangeable.