BMC Cardiovascular Disorders (Dec 2023)
Loss of capture of conduction system pacemaker caused by fibrosis surrounding the lead: a case report
Abstract
Abstract Background Conduction system pacing (CSP) is a novel technique that involves pacing the His-Purkinje system instead of the traditional right ventricular (RV) apex. This technique aims to avoid the adverse effects of RV apical pacing, which can lead to ventricular dyssynchrony and heart failure over time. CSP is gaining popularity but its long-term efficacy and challenges remain uncertain. This report discusses a case where CSP was initially successful but faced complications due to an increasing pacing threshold. Case presentation A 65-year-old female with total atrioventricular block was referred for brady-pacing. Due to the potential for chronic RV pacing, CSP was chosen. The CSP implantation involved subcutaneous device placement, with a CSP lead in the left bundle branch area (LBBA) and an RV backup lead. A year after successful implantation, the LBBA pacing threshold progressively increased. Subsequent efforts to correct it led to anodal capture and battery depletion. Cardiac magnetic resonance imaging (CMR) revealed mid-septal fibrosis at the area of LBBA lead placement and suggested cardiac sarcoidosis as a possible cause. Conclusion CSP is a promising technique for treating bradyarrhythmias, but this case underscores the need for vigilance in monitoring pacing thresholds. Increasing thresholds can render CSP ineffective, necessitating alternative pacing methods. The CMR findings of mid-septal fibrosis and the potential diagnosis of cardiac sarcoidosis emphasize the importance of pre-implantation assessment, as CSP may be compromised by underlying structural abnormalities. This report highlights the complexities of pacing strategy selection and the significance of comprehensive evaluation before adopting CSP.
Keywords