Scientific Reports (May 2024)
Quantitative analysis of blooming artifact caused by calcification based on X-ray energy difference using computed tomography
Abstract
Abstract Blooming artifacts caused by calcifications appearing on computed tomography (CT) images lead to an underestimation of the coronary artery lumen size, and higher X-ray energy levels are suggested to reduce the blooming artifacts with subjective visual assessment. This study aimed to evaluate the effect of higher X-ray energy levels on the quantitative measurement of adjacent pixels affected by calcification using CT images. In this two-part study, CT images were acquired from dual-energy CT scanners by changing the X-ray energy levels such as kilovoltage peak (kVp) and kilo-electron volts (keV). Adjacent pixels affected by calcification were measured using the brightened length, excluding the actual calcified length, as determined by the full width at third maximum. In a separate clinical study, the adjacent affected pixels associated with 23 calcifications across 10 patients were measured using the same method as that used in the phantom study. Phantom and clinical studies showed that the change in kVp (field of view [FOV] 300 mm: p = 0.167, 0.494, and 0.861 for vendors 1, 2, and 3, respectively) and keV levels (p = 0.178 for vendor 2) failed to reduce the adjacent pixels affected by calcification, respectively. Moreover, the change in keV levels showed different aspects of adjacent pixels affected by calcification in the phantom study (FOV 300 mm: no significant difference [p = 0.191], increase [p < 0.001], and decrease [p < 0.001] for vendors 1, 2, and 3, respectively). Quantitative measurements revealed no significant relationship between higher X-ray energy levels and the adjacent pixels affected by calcification.