Physics and Imaging in Radiation Oncology (Apr 2024)

Clinical implementation of a commercial synthetic computed tomography solution for radiotherapy treatment of glioblastoma

  • Sevgi Emin,
  • Elia Rossi,
  • Elisabeth Myrvold Rooth,
  • Torsten Dorniok,
  • Mattias Hedman,
  • Giovanna Gagliardi,
  • Fernanda Villegas

Journal volume & issue
Vol. 30
p. 100589

Abstract

Read online

Background and Purpose: Magnetic resonance (MR)-only radiotherapy (RT) workflow eliminates uncertainties due to computed tomography (CT)-MR image registration, by using synthetic CT (sCT) images generated from MR. This study describes the clinical implementation process, from retrospective commissioning to prospective validation stage of a commercial artificial intelligence (AI)-based sCT product. Evaluation of the dosimetric performance of the sCT is presented, with emphasis on the impact of voxel size differences between image modalities. Materials and methods: sCT performance was assessed in glioblastoma RT planning. Dose differences for 30 patients in both commissioning and validation cohorts were calculated at various dose-volume-histogram (DVH) points for target and organs-at-risk (OAR). A gamma analysis was conducted on regridded image plans. Quality assurance (QA) guidelines were established based on commissioning phase results. Results: Mean dose difference to target structures was found to be within ± 0.7 % regardless of image resolution and cohort. OARs’ mean dose differences were within ± 1.3 % for plans calculated on regridded images for both cohorts, while differences were higher for plans with original voxel size, reaching up to −4.2 % for chiasma D2% in the commissioning cohort. Gamma passing rates for the brain structure using the criteria 1 %/1mm, 2 %/2mm and 3 %/3mm were 93.6 %/99.8 %/100 % and 96.6 %/99.9 %/100 % for commissioning and validation cohorts, respectively. Conclusions: Dosimetric outcomes in both commissioning and validation stages confirmed sCT’s equivalence to CT. The large patient cohort in this study aided in establishing a robust QA program for the MR-only workflow, now applied in glioblastoma RT at our center.

Keywords