Diagnostics (Aug 2022)
Is It Possible to Replace Conventional Radiography (CR) with a Dose Neutral Computed Tomography (CT) of the Cervical Spine in Emergency Radiology—An Experimental Cadaver Study
Abstract
The purpose of this experimental study on recently deceased human cadavers was to investigate whether (I) the radiation exposure of the cervical spine CT can be reduced comparable to a dose level of conventional radiography (CR); and (II) whether and which human body parameters can be predictive for higher dose reduction potential (in this context). Materials and Methods: Seventy serial CT scans of the cervical spine of 10 human cadavers undergoing postmortem virtual autopsy were taken using stepwise decreasing upper limits of the tube current (300 mAs, 150 mAs, 110 mAs, 80 mAs, 60 mAs, 40 mAs, and 20 mAs) at 120 kVp. An additional scan acquired at a fixed tube current of 300 mAs served as a reference. Images were reconstructed with filtered back projection and the upper (C1-4) and lower (C4-7) cervical spine were evaluated by three blinded readers for image quality, regarding diagnostic value and resolution of anatomical structures according to a semiquantitative three-point-scale. Dose values and individual physical parameters were recorded. The relationship of diagnostic IQ, dose reduction level, and patients’ physical parameters were investigated. The high-contrast resolution of the applied CT protocols was tested in an additional phantom study. Results: The IQ of the upper cervical spine was diagnostic at 1.69 ± 0.58 mGy (CTDI) corresponding to 0.20 ± 0.07 mSv (effective dose) in all cadavers. IQ of the lower cervical spine was diagnostic at 4.77 ± 1.86 mGy corresponding to 0.560 ± 0.21 mSv (effective dose) in seven cadavers and at 2.60 ± 0.93 mGy corresponding to 0.31 ± 0.11 mSv in four cadavers. Significant correlation was detected for BMI (0.8366; p = 0.002548) and the anteroposterior (a.p.) chest diameter (0.8363; p = 0.002566), shoulder positioning (0.79799; p = 0.00995), and radiation exposure. Conclusions: Conventional radiography can be replaced with a nearly dose-neutral CT scan of the cervical spine.
Keywords