Trends in Hearing (Feb 2019)
Improving Localization and Speech Reception in Noise for Bilateral Cochlear Implant Recipients
Abstract
This study looked at different methods to preserve interaural level difference (ILD) cues for bilateral cochlear implant (BiCI) recipients. One possible distortion to ILD is from automatic gain control (AGC). Localization accuracy of BiCI recipients using default versus increased AGC threshold and linked AGCs versus independent AGCs was examined. In addition, speech reception in noise was assessed using linked versus independent AGCs and enabling and disabling Autosensitivity™ Control. Subjective information via a diary and questionnaire was also collected about maps with linked and independent AGCs during a take-home experience. Localization accuracy improved in the increased AGC threshold and the linked AGCs conditions. Increasing the AGC threshold resulted in a 4° improvement in root mean square error averaged across all speaker locations. Using linked AGCs, BiCI participants experienced an 8° improvement for all speaker locations and a 19° improvement at the speaker location most affected by the AGC. Speech reception threshold in noise improved by an average of 2.5 dB when using linked AGCs versus independent AGCs. In addition, the effect of linked AGCs on speech in noise was compared with that of Autosensitivity™ Control. The Speech, Spatial, and Qualities of Hearing Scale-12 question comparative survey showed an improvement when using maps with linked AGCs. These findings support the hypothesis that ILD cues may be preserved by increasing the AGC threshold or linking AGCs.