Frontiers in Endocrinology (Nov 2023)

Feasibility of parathyroid gland autofluorescence imaging after indocyanine green fluorescence angiography

  • Marieke Richard,
  • Philippe Rizo

DOI
https://doi.org/10.3389/fendo.2023.1248449
Journal volume & issue
Vol. 14

Abstract

Read online

BackgroundIn thyroid surgery, autofluorescence allows the parathyroid glands (PTGs) to be located very early to protect them. Moreover, indocyanine green (ICG) fluorescence angiography (ICG-FA) allows for assessing the viability of the PTGs and identifying blood vessels to preserve them. The main limitation of using ICG-FA is that once ICG has been injected, it is no longer possible to observe PTG autofluorescence using existing devices. This study aimed to describe an approach that allows for visualization of the PTGs by autofluorescence, even after ICG injection.MethodsWe redesigned the FLUOBEAM® LX system to excite fluorescence at 685 nm and detect fluorescence between 700 and 900 nm. This device had short-pass filters at 775 nm that helped to split the contributions of the PTG autofluorescence and ICG fluorescence. Tests were performed on extemporaneous PTG preparations placed next to ICG droplets to assess for rejection of the ICG signal.ResultsA low-pass filter at 775 nm detected 60% of the autofluorescence signals and 10% of the ICG signals.ConclusionThese findings support the possibility of visualizing PTG autofluorescence despite multiple ICG injections and measuring the balance between ICG and autofluorescence signals.