This report discusses a case of transient 2:1 atrioventricular block with conduction system pacing 4 hours after leadless right ventricular pacemaker implantation in a 19-year-old patient with a history of cardioinhibitory syncope and asystole cardiac arrest but without preexisting atrioventricular block. The atrioventricular block was resolved spontaneously. Pacing morphology was suggestive of right bundle branch pacing. Neither 2:1 atrioventricular block nor conduction system pacing has previously been a reported outcome of right ventricular leadless pacemaker implantation. The report demonstrates that conduction system pacing with leadless devices is achievable. Further study of techniques, limitations, and complications related to intentional right ventricular leadless conduction system pacing is warranted.