Applied Sciences (Feb 2023)

Static Bacterial Leakage in Different Conometric Connections: An In Vitro Study

  • Simonetta D’Ercole,
  • Tatiane Cristina Dotta,
  • Giovanna Iezzi,
  • Alessandro Cipollina,
  • Vinicius Pedrazzi,
  • Adriano Piattelli,
  • Morena Petrini

DOI
https://doi.org/10.3390/app13042693
Journal volume & issue
Vol. 13, no. 4
p. 2693

Abstract

Read online

This in vitro study aims to evaluate the bacterial microleakage of three conometric connections. Sixty dental implants (3P implafavourite Scalenghe) were divided in groups (n = 20): Cone–Morse with passing screw (Group 1); Cone–Morse with solid abutment (Group 2); and Conometric connection with esthetic abutment (Group 3). The implants were fixed in resin bases. Then, 1.0 µL of Streptococcus oralis (SO) was inoculated in the internal platform in 10 fixtures for each group, and another 10 were inoculated with Pseudomonas aeruginosa (PA). The abutments were then screwed, and five implants from each subgroup were randomly selected for SEM inspection to ensure that the abutments were installed correctly. Data were submitted to statistical analysis, ANOVA and Fisher’s Least Significant Difference (p ≤ 0.05). The turbidity of the broth was monitored for 14 days of follow-up in order to determine the penetration of the bacterial suspension into the surrounding solution, but the observation of the samples lasted until the 90th day, in which there was no difference between the two. Microbial contamination was found in 30%, 20%, and 50% of Group 1, Group 2, and Group 3, respectively, but there were no statistically significant differences between the groups, and PA showed greater infiltration than SO. Although no statistically significant differences were found, cone morse connections showed lower infiltration percentages, respective to the conometric connection with 18° angle.

Keywords