Journal of Medical Physics (Jan 2012)

Patient dose analysis in total body irradiation through in vivo dosimetry

  • K Ganapathy,
  • P. G. G. Kurup,
  • V Murali,
  • M Muthukumaran,
  • N Bhuvaneshwari,
  • J Velmurugan

DOI
https://doi.org/10.4103/0971-6203.103607
Journal volume & issue
Vol. 37, no. 4
pp. 214 – 218

Abstract

Read online

Total body irradiation (TBI) is a special radiotherapy technique, administered prior to bone marrow transplantation. Due to the complex nature of the treatment setup, in vivo dosimetry for TBI is mandatory to ensure proper delivery of the intended radiation dose throughout the body. Lithium fluoride (LiF) TLD-100 chips are used for the TBI in vivo dosimetry. Results obtained from the in vivo dosimetry of 20 patients are analyzed. Results obtained from forehead, abdomen, pelvis, and mediastinum showed a similar pattern with the average measured dose from 96 to 97% of the prescription dose. Extremities and chest received a dose greater than the prescription dose in many instances (more than 20% of measurements). Homogeneous dose delivery to the whole body is checked by calculating the mean dose with standard deviation for each fraction. Reasons for the difference between prescription dose and measured dose for each site are discussed. Dose homogeneity within ±10% is achieved using our in-house TBI protocol.