Behavioral Sciences (May 2021)

Alternative Devices for Heart Rate Variability Measures: A Comparative Test–Retest Reliability Study

  • Jacquelin M. Killian,
  • Rachel M. Radin,
  • Cubby L. Gardner,
  • Lalon Kasuske,
  • Kylee Bashirelahi,
  • Dominic Nathan,
  • David O. Keyser,
  • Christopher J. Cellucci,
  • David Darmon,
  • Paul E. Rapp

DOI
https://doi.org/10.3390/bs11050068
Journal volume & issue
Vol. 11, no. 5
p. 68

Abstract

Read online

Using healthy adult participants, seven measures of heart rate variability were obtained simultaneously from four devices in five behavioral conditions. Two devices were ECG-based and two utilized photoplethysmography. The 140 numerical values (measure, condition, device) are presented. The comparative operational reliability of the four devices was assessed, and it was found that the two ECG-base devices were more reliable than the photoplethysmographic devices. The interchangeability of devices was assessed by determining the between-device Limits of Agreement. Intraclass correlation coefficients were determined and used to calculate the standard error of measurement and the Minimal Detectable Difference. The Minimal Detectable Difference, MDD, quantifies the smallest statistically significant change in a measure and is therefore critical when HRV measures are used longitudinally to assess treatment response or disease progression.