Journal of Innovative Optical Health Sciences (Jul 2022)

Implementation and uniformity calibration of LED array for photodynamic therapy

  • Xu Wang,
  • Jiyuan Xiong,
  • Xiaoming Hu,
  • Qin Li

DOI
https://doi.org/10.1142/S1793545822400041
Journal volume & issue
Vol. 15, no. 04

Abstract

Read online

Irradiance uniformity is an important parameter for a Photodynamic Therapy (PDT) device. The calibration and verification of a LED array light source based on computer vision technology were implemented and carried out. First, the correlation coefficients between the pulse width modulate value and the irradiance were calibrated. Then, the correction of the actual light center and divergence angle were solved by image processing to reduce errors from each LED lens. Finally, uniformity was optimized according to the irradiance formula of the Lambertian source. The lowest coefficients of variation of irradiance were 4.87% in a [Formula: see text] area and 3.55% in a [Formula: see text] area within the depth range of 8–12[Formula: see text]cm when the expected irradiance was 100[Formula: see text]mW/cm2. This finding indicated that the light source can achieve a more uniform illumination and provide a better therapeutic effect for the PDT of port-wine stains.

Keywords