Technology in Cancer Research & Treatment (Sep 2023)

Nondeformed Ultrasound Image Production Method for Ultrasound-Guided Radiotherapy

  • Qixuan Li MS,
  • Fan Zhang MS,
  • Qianyi Xi MS,
  • Zhuqing Jiao PhD,
  • Xinye Ni PhD

DOI
https://doi.org/10.1177/15330338231194546
Journal volume & issue
Vol. 22

Abstract

Read online

Purpose: During ultrasound (US)-guided radiotherapy, the tissue is deformed by probe pressure, and the US image is limited by changes in tissue and organ position and geometry when the US image is aligned with computed tomography (CT) image, leading to poor alignment. Accordingly, a pixel displacement-based nondeformed US image production method is proposed. Methods: The correction of US image deformation is achieved by calculating the pixel displacement of an image. The positioning CT image (CT std ) is used as the gold standard. The deformed US image (US def ) is inputted into the Harris algorithm to extract corner points for selecting feature points, and the displacement of adjacent pixels of feature points in the US video stream is calculated using the Lucas-Kanade optical flow algorithm. The moving least squares algorithm is used to correct US def globally and locally in accordance with image pixel displacement to generate a nondeformed US image (US rev ). In addition, US def and US rev were separately aligned with CT std to evaluate the improvement of alignment accuracy through deformation correction. Results: In the phantom experiment, the overall and local average correction errors of the US image under the optimal probe pressure were 1.0944 and 0.7388 mm, respectively, and the registration accuracy of US def and US rev with CT std was 0.6764 and 0.9016, respectively. During the volunteer experiment, the correction error of all 12 patients’ data ranged from −1.7525 to 1.5685 mm, with a mean absolute error of 0.8612 mm. The improvement range of US and CT registration accuracy, before and after image deformation correction in the 12 patients evaluated by a normalized correlation coefficient, was 0.1232 to 0.2476. Conclusion: The pixel displacement-based deformation correction method can solve the limitation imposed by image deformation on image alignment in US-guided radiotherapy. Compared with US def , the alignment results of US rev with CT were better.