The Egyptian Journal of Radiology and Nuclear Medicine (Apr 2020)
Role of shear wave sono-elastography (SWE) in characterization of hepatic focal lesions
Abstract
Abstract Background Focal liver lesions are considered a major problem during abdominal examinations. Shear wave sono-elastography (SWE) has been demonstrated to be helpful in assessment of liver fibrosis degree. The purpose of this study was to evaluate the role of SWE in characterization of benign and malignant hepatic focal lesions. Results Seventy-five (75) patients with variable focal liver lesions (52 malignant and 23 benign) were analyzed by SWE. The stiffness values of surrounding hepatic parenchyma were also measured as a reference for readings of the focal lesion stiffness values. Final diagnosis was achieved by core needle biopsy (in 1 benign and 38 malignant cases) and contrast enhanced CT and MRI (in all cases). Cholangiocarcinoma (CCC) was the stiffest malignant lesion with median stiffness value (35.9 kPa). Focal nodular hyperplasia (FNH) was the stiffest benign lesion (26.7 kPa). The median stiffness value of malignant focal lesions (20.22 kPa) was significantly higher than that of benign focal lesions (10.68 kPa) (P value < 0.001). ROC curve of SWE median stiffness values for differentiation of benign from malignant hepatic focal lesions had AUC = 0.834, and using cut of value 14.165 kPa, yielding 98.1% sensitivity, 78.3% specificity, and 92% accuracy. Conclusion SWE has high accuracy in differentiating benign form malignant liver focal lesions with promising results in individual characterization of some malignant (HCC and CCC) and benign hepatic focal lesion (FNH from other benign lesions).